| Parameter | Specification | ||||||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Model Series | CDLT Series Ultra-Low Temperature Deep Cold Test Chamber | ||||||||||||||||||||||||||||||||||||||||||
| Dimensional & Power Models |
|
||||||||||||||||||||||||||||||||||||||||||
| Performance Parameters |
• Temperature Range: -80°C ~ +130°C (Optional: -90°C ~ -120°C / -200°C ultra-low) • Temperature Fluctuation: ≤ ±0.5°C • Temperature Uniformity: ≤ ±2.0°C • Temperature Deviation: ≤ ±2.0°C • Heating Rate: -80°C → +130°C ≤ 80 min (Empty load) • Cooling Rate: +20°C → -80°C ≤ 80 min (Empty load; up to 40°C/min max rate) |
||||||||||||||||||||||||||||||||||||||||||
| Chamber Structure & Materials |
• Inner Liner: 1.5mm thick stainless steel • Outer Shell: 1.5mm cold-rolled steel with baking paint • Insulation Layer: 100mm thick polyurethane (-80°C~0°C) / 150mm thick polyurethane (-80°C~-120°C) |
||||||||||||||||||||||||||||||||||||||||||
| Refrigeration System |
• Refrigeration Mode: Mechanical compressor cascade / multi-stage parallel system (Air-cooled / Water-cooled) • Compressor: Fully hermetic compressor • Refrigerant: Mixed liquid refrigerant / Liquid Nitrogen dynamic control • Cooling Method: Air cooling / Water cooling |
||||||||||||||||||||||||||||||||||||||||||
| Control System |
• Temperature Resolution: ±0.1°C • Controller: 7.0" true color touchscreen, Chinese/English input, history curve logging, PID self-tuning, programmable cycle control • Temperature Sensor: Platinum resistance PT100 armored high-precision sensor |
||||||||||||||||||||||||||||||||||||||||||
| Heating System | Nichrome alloy electric heater, finned heating tube | ||||||||||||||||||||||||||||||||||||||||||
| Standard Accessories | Observation window ×1 set, Sample shelves ×2 pcs, Cable port Ø50mm ×1 pc | ||||||||||||||||||||||||||||||||||||||||||
| Safety Protection | Leakage protection, short circuit protection, over-temperature protection, motor overheating, compressor overpressure, overload, and overcurrent protection | ||||||||||||||||||||||||||||||||||||||||||
| Power Supply | AC 380V 50Hz, 3-Phase 5-Wire System | ||||||||||||||||||||||||||||||||||||||||||
| Compliance Standards |
|
||||||||||||||||||||||||||||||||||||||||||
| Test Constraints / Prohibitions |
• Testing and storage of flammable, explosive, or highly volatile substance samples are strictly prohibited. • Testing and storage of corrosive substance samples are strictly prohibited. • Testing or storage of biological samples is strictly prohibited. • Testing and storage of strong electromagnetic emission source samples are strictly prohibited. |
||||||||||||||||||||||||||||||||||||||||||
| Operating Environment |
• Temperature: 5°C ~ 25°C, Humidity: ≤ 85% RH. • No strong vibration; sufficient ventilation and maintenance clearance around walls/equipment. • Free of high concentration dust and corrosive substances; equipped with floor drainage near chamber. |
||||||||||||||||||||||||||||||||||||||||||
| After-Sales Service |
• 1-Year Warranty & Lifetime Maintenance; customer service profile setup with regular follow-ups. • Free lifetime technical support. Phone response within 1 hour; field dispatch within 24 hours if needed. • Free delivery, installation, commissioning, and hands-on operational training for technical personnel. |